This sensor is appropriate for the monitoring of the intensity of UV radiation coming from UV lamps:
It's main features are:
GaP-based sensor; compatible with old sensor types on the market because of the higher output current but please note the sensitivity range below
no problem for using with low pressure lamps, for medium pressure lamps and measurements affected by daylight we recommend SUV sensor types (daylight blind)
Probe body is 0 20 x 33 mm, made of 1.4404 stainless steel, with external G 1/4" SW 17 thread
Entry window for UV: 0 6.0 mm; angle of opening 30°
Operating temperature: 0 to 60°C; 100°C for brief periods, storage temperature: -40 to 70°C
Option A1: Sealed and watertight with a window permeable for UV radiation,
pressure resistant up to 16 bar overpressure and down to 1 bar sub-atmospheric pressure
Option C (is standard): Series 763 flanged plug, watertight to IP65 when screwed in, matching Z1 G connector cable (5 m of 5-pole cable with open end)